Product Description
Bet Analyzer is an instrument used to measure the specific surface area and pore structure of porous materials. It calculates the specific surface area and pore size distribution by gas adsorption method and is mainly used for material characterization and performance evaluation.
Advantages
Strong Micropore Resolution
The vacuum limit reaches 1×10⁻⁶ Pa. Combined with a highly sensitive sensor, it can accurately measure pores as small as 0.35 nm, making it ideal for microporous material analysis.
High Sample Throughput
Supports up to three analysis stations and six external degassing stations, enabling simultaneous processing of multiple samples.
High Gas Control Precision
A built-in PID airflow control system prevents data interference from airflow fluctuations.
Long-Term Low-Intervention Operation
Equipped with a maximum 4L Dewar flask, it offers a holding time of up to 100 hours, enabling continuous operation for multiple test cycles.
Strong Low-Adsorption Detection Capability
Supports sample volumes as low as 0.1 mL, making it suitable for precise surface area analysis of high-value or very small samples.
Strong Data Compatibility
Raw data can be directly exported to TXT format for easy import into plotting tools such as Origin.
Specifications
|
Control Mode |
Automatic controlled by software |
|
Communication mode |
RS485, upgraded to the USB or network interface |
|
Built-in Detector |
Three pressure sensors, including a pressure sensor of pirani vacuum silicon |
|
Analysis station |
1~3 |
|
Volume of sampling |
0.1-3mL |
|
Reaction Gas |
N2, Ar,CO2 |
|
Gas Purity |
>99.995 % |
|
Calibration Gas |
He |
|
Control of Gas Flow |
PID |
|
Sample Cell |
Erect type, quartz material, half- round bottom |
|
Degassing Station |
1-3 degassing stations of built-in,6 degassing stations of external sample cell. |
|
All in One Valve Unit |
Integrated valve bank, pile-up valve, gas circuit, pressure sensor |
|
Vacuum System |
Two-stage rotary vacuum pump, turbo molecular pump if necessary |
|
Vacuum Purity |
1*10-2 pa, 1*10-6 pa can be reached if on the condition of molecular pump |
|
Dewar |
2L-4L high capacity, staying warm is no more than 100 hours |
|
Jacking System |
Automatically controlled by lift |
|
Software |
Dedicated analysis workstation of software,based on Windows platform software |
|
Data Export |
Support TXT text output,which makes the exchange data with drawing tools like Origin more convenient |
FAQ
Q: What is the minimum sample volume that the Bet Analyzer can measure?
A: It supports sample volumes as low as 0.1 mL, making it suitable for accurate testing of small samples.
Q: Can multiple samples be tested simultaneously?
A: It supports up to three analysis stations and six degassing positions, enabling parallel processing of multiple samples.
Q: Is it suitable for testing microporous materials?
A: Equipped with a high vacuum system, it can achieve a vacuum level of up to 1×10⁻⁶ Pa, making it suitable for measuring micropores smaller than 1 nm.
Q: Is the gas flow rate stable during the test?
A: PID gas flow control ensures smooth airflow during the adsorption process and highly consistent data.
Q: Can test data be exported for secondary analysis?
A: Data can be exported in TXT format, allowing direct use in software such as Origin for charting and calculations.
Hot Tags: bet analyzer, China bet analyzer manufacturers, suppliers, factory, viscometer, fluorometer, cooler, high pressure submersible pump, turbidity meter, low noise submersible pump